微信公众号随时随地查标准

QQ交流1群(已满)

QQ群标准在线咨询2

QQ交流2群

购买标准后,可去我的标准下载或阅读

4.1 This practice is used to outline the minimum necessary elements and conditions to obtain an accurate determination of the location of wet insulation in roofing systems using infrared imaging.4.2 This practice is not meant to be an instructional document or to provide all the knowledge and background necessary to provide an accurate analysis. For further information, see ANSI-ASHRAE Standard 101 and ISO/DP 6781.3E.4.3 This practice does not provide methods to determine the cause of moisture or its point of entry. It does not address the suitability of any particular system to function capably as waterproofing.1.1 This practice applies to techniques that employ infrared imaging at night to determine the location of wet insulation in roofing systems that have insulation above the deck in contact with the waterproofing. This practice includes ground-based and aerial inspections. (Warning—Extreme caution shall be taken when accessing or walking on roof surfaces and when operating aircraft at low altitudes, especially at night.) (Warning—It is a good safety practice for at least two people to be present on the roof surface at all times when ground-based inspections are being conducted.)1.2 This practice addresses criteria for infrared equipment such as minimum resolvable temperature difference, spectral range, instantaneous field of view, and field of view.1.3 This practice addresses meteorological conditions under which infrared inspections shall be performed.1.4 This practice addresses the effect of roof construction, material differences, and roof conditions on infrared inspections.1.5 This practice addresses operating procedures, operator qualifications, and operating practices.1.6 This practice also addresses verification of infrared data using invasive test methods.1.7 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in 1.1.1.9 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

定价: 590元 / 折扣价: 502 加购物车

在线阅读 收 藏

4.1 The test results provide an indication of the turbine-powered nozzle life. The end of turbine life will be judged in accordance with 3.1.1.1.1 This test method covers the turbine-powered nozzle used in household central vacuum cleaning systems.1.2 This test method provides a test for determining the operating turbine life in hours by an accelerated laboratory procedure. The turbine is tested while mounted and operated in the power nozzle.1.3 This test method covers only the turbine-powered nozzle. The system used to provide the airflow source is not under consideration.1.4 This test method is limited to the determination of turbine life for a household turbine-powered nozzle.1.5 The values stated in inch-pound units are to be regarded as the standard. The values given in parentheses are for information only.1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

定价: 590元 / 折扣价: 502 加购物车

在线阅读 收 藏

5.1 This test method is used by athletic footwear manufacturers and others, both as a tool for development of athletic shoe cushioning systems and as a test of the general cushioning characteristics of athletic footwear products, materials and components. Adherence to the requirements and recommendations of this test method will provide repeatable results that can be compared among laboratories.5.2 Data obtained by these procedures are indicative of the impact attenuation of athletic shoe cushioning systems under the specific conditions employed.5.3 This test method is designed to provide data on the force versus displacement response of athletic footwear cushioning systems under essentially uniaxial impact loads at rates that are similar to those of heel and forefoot impacts during different athletic activities.5.4 The peak or maximum values of force, acceleration, displacement, and strain are dependent on the total impact energy applied to the specimen. These values are normalized to provide comparative results for a reference value of total energy input.5.5 Impact attenuation outcomes are strongly dependent on initial conditions (impact mass, impact velocity, contact area, etc.) and on specimen size and the specimen’s prior history of compressive loading. Therefore results should be compared only for specimens of the same nominal size and prior conditioning.Note 1—Impact test outcomes have been found to correlate with in-vivo loads (peak ground reaction force, peak plantar pressure, lower extremity acceleration) experienced by runners. Relationships between test outcomes and subjective perceptions of cushioning have also been found. However, there is no direct evidence of a correlation between scores on this test method and the probability of injury among users of a particular athletic footwear product.1.1 This test method describes the use of a gravity-driven impact test to measure certain impact attenuation characteristics of cushioning systems and cushioning materials employed in the soles of athletic shoes.1.2 This test method uses an 8.5 kg mass dropped from a height of 30-70 mm to generate force-time profiles that are comparable to those observed during heel and forefoot impacts during walking, running and jump landings.1.3 This test method is intended for use on the heel and or forefoot regions of whole, intact athletic shoe cushioning systems. An athletic shoe cushioning system is defined as all of the layers of material between the wearer's foot and the ground surface that are normally considered a part of the shoe. This may include any of the following components: outsole or other abrasion resistant outer layer, a midsole of compliant cushioning materials or structures forming an intermediate layer, an insole, insole board, or other material layers overlying the midsole, parts of the upper and heel counter reinforcement which extend beneath the foot, and an insock, sockliner or other cushioning layers, either fixed or removable, inside the shoe.1.4 This test method may also be employed in to measure the impact attenuation of cushioning system components and cushioning material specimens.1.5 This test method is not intended for use as a test of shoes classified by the manufacturer as children's shoes.1.6 The type, size or dimensions and thickness of the specimen, the total energy input and prior conditioning shall qualify test results obtained by this test method.1.6.1 The range of tests results is limited by the calibrated range of the test device’s force transducer. Combinations of thin specimens, high specimen stiffness and high total energy input may produce forces that exceed the transducer’s capacity and are hence not measurable. In practice, the specified force transducer range (10 kN) accommodates more than 99 % of typical shoe soles and cushioning material specimens that are 7 mm or more in thickness at a total energy input of 5 Joules.1.6.2 The nominal value of the total energy input applied by this test method is 5 J for shoes, such as running shoes, which are subject to moderate impacts during normal use. Total energy inputs of 7.0 J and 3.0 J may be used for shoes (e.g basketball shoes) which are subject to higher impact loads during normal use. Other values of total energy input may be used, if they are stated in the report.1.6.3 Results from tests performed with different total energy inputs or with different masses are not directly comparable.1.6.4 Specimen thickness has a significant effect on impact attenuation outcomes. Consequently, results from tests of material specimens of different thicknesses cannot be directly compared.1.6.5 The impact attenuation of cushioning materials may change over time and with use (e.g. wear or durability testing) or prior conditioning (e.g. from previous tests). Consequently, test results obtained using this test method shall be qualified by the age and prior conditioning of the samples.1.7 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

定价: 0元 / 折扣价: 0

在线阅读 收 藏

4.1 This practice is to be used as a guide to classify water bodies for spill control systems. These classifications may be used in formulating standards for design, performance, evaluation, contingency and response planning, contingency and response plan evaluation, and standard practice for spill control systems.4.2 Relatively few parameters of broad range have been used in Table 1 in order to enable the user to readily identify general conditions under which spill control systems can be used.4.3 Satisfactory operation of any specific spill control systems may not extend over the full range of conditions identified by Table 1. Detailed discussion with systems suppliers is recommended.4.4 Effective operation of oil spill control equipment depends on many factors, of which the prevailing environmental conditions are just a few. Factors such as, but not limited to, deployment techniques, level of training, personnel performance, and mechanical reliability can also affect equipment performance.1.1 This practice creates a system of categories that classify water bodies relating to the control of spills of oil and other substances that float on or into a body of water.1.2 This practice does not address the compatibility of spill control equipment with spill products. It is the user's responsibility to ensure that any equipment selected is compatible with anticipated products.1.3 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

定价: 515元 / 折扣价: 438 加购物车

在线阅读 收 藏

This specification covers aluminum and aluminum-alloy seamless pipe and seamless extruded tube for gas and oil transmission and distribution piping systems. The pipe and tube shall be produced from hollow extrusion ingot (cast in hollow form or pierced) and shall be extruded by use of the die and mandrel method. The pipe and tube shall conform to the chemical composition requirements specified. The determination of chemical composition shall be made in accordance with suitable chemical (test methods E 34), or spectrochemical (test methods E 227, E 607, and E 1251) methods. Heat treatment for the production of T1 and T5-type tempers shall be in accordance with Practice B 807, and for the production of T4 and T6-type tempers, except as noted, shall be in accordance with practice B 918. Unless otherwise specified, alloys 6061, 6063, and 6351 may be solution heat treated and quenched at the extrusion press in accordance with practice B 807 for the production of T4 and T6-type tempers, as applicable. The material shall conform to the tensile property requirements specified. The tension tests shall be made in accordance with test methods B 557 and B 557M. Pipe and tube heat treated at the extrusion press shall conform to all requirements specified.1.1 This specification covers seamless pipe and seamless extruded tube in the aluminum and aluminum alloys (Note 1) and tempers listed in Table 1 and Table 2, respectively. Seamless pipe and seamless tube are intended for use in applications involving internal pressure.Note 1—Throughout this specification use of the term alloy in the general sense includes aluminum as well as aluminum alloy.Note 2—For drawn seamless tubes, see Specifications B210 and B210M; for extruded tubes, Specifications B221 and B221M; for drawn seamless tubes for condensers and heat exchangers, Specifications B234 and B234M; for seamless pipe and seamless extruded tube, B241/B241M; for round welded tubes, Specification B313/B313M; for seamless condenser and heat exchanger tubes with integral fins, Specification ; for extruded structural pipe and tube, Specification B429/B429M; and for drawn tube for general purpose applications, Specification B483/B483M.1.2 Alloy and temper designations are in accordance with ANSI H35.1 [H35.1M]. The equivalent Unified Numbering System alloy designations are those of Table 3 preceded by A9, for example, A93003 for aluminum alloy 3003 in accordance with Practice E527.1.3 For acceptance criteria for inclusion of new aluminum and aluminum alloys in this specification, see Annex A2.1.4 The values stated in either inch-pound units or SI units are to be regarded separately as standard. Within the text, the SI units are shown in brackets. The values stated in each system are not exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with the specification.TABLE 1 Tensile Property Limits for Extruded Seamless PipeA,BAlloy Temper Pipe Size,in. Strength, min, ksi [MPa] ElongationC,DTensile Yield (0.2 % Offset) in 2 in. [50 mm] or 4×Diameter, min, % in 5 × D(5.65)3003 H18 under 1 27.0 [185] 24.0 [165] 4 4 H112 1 and over 14.0 [95] 5.0 [35] 25 226061 T6 under 1 38.0 [260] 35.0 [240] 8 ... 1 and over 38.0 [260] 35.0 [240] 10E 96063 T6 all 30.0 [205] 25.0 [170] 8 76351 T5T6 allall 38.0 [260]42.0 [290] 35.0 [240]37.0 [255] 10E10F 99A The basis for establishment of mechanical property limits is given in Annex A1 of this specification.B To determine conformance to this specification, each value for tensile strength and for yield strength shall be rounded to the nearest 0.1 ksi [MPa] and each value for elongation to the nearest 0.5 %, both in accordance with the rounding method of Practice E29.C Elongation of full-section and sheet-type specimens is measured in 2 in.; of cut-out round specimens, 4× specimen diameter.D Elongations in 50 mm apply for pipe tested in full sections and for sheet-type specimens machined from material up through 12.5 mm in thickness having parallel surfaces. Elongations in 5 × D (at 5.65), where D and A are diameter and cross-sectional area of the specimen, respectively, apply to round test specimens machined from thicknesses over 6.30 mm.E The minimum elongation for a wall thickness up through 0.249 in. [6.3 mm] is 8 %.F For wall thickness 0.124 in. [3.20 mm] and less, the minimum elongation is 8 %.TABLE 2 Tensile Property Limits for Extruded Seamless TubeA,BTemper Specified WallThickness, in. [mm] Area, in.2 [mm2] Tensile Strength, ksi [MPa] Yield Strength(0.2 % offset)ksi [MPa], min ElongationC,Dmin max in 2 in. [50 mm] or4 × D min,% in 5 × D(5.65)EAluminum 1060FOH112 allall allall 8.5 [60]8.5 [60] 14.0 [95]... [...] 2.5 [15]2.5 [15] 2525G 2222GAlloy 3003FOH112 allall allall 14.0 [95]14.0 [95] 19.0 [130]... [...] 5.0 [35]5.0 [35] 2525 2222Alloy Alclad 3003FOH112 allall allall 13.0 [90]13.0 [90] 18.0 [125]... [...] 4.5 [30]4.5 [30] 2525 2222Alloy 5083FOH111H112 all [130.00]all [130.00]all [130.00] up through 32.0 [20 000]up through 32.0 [20 000]up through 32.0 [20 000] 39.0 [270]40.0 [275]39.0 [270] 51.0 [350]... [...]... [...] 16.0 [110]24.0 [165]16.0 [110] 141212 121010Alloy 5086FOH111H112 all [130.00]all [130.00]all [130.00] up through 32.0 [20 000]up through 32.0 [20 000]up through 32.0 [20 000] 35.0 [240]36.0 [250]35.0 [240] 46.0 [315]... [...]... [...] 14.0 [95]21.0 [145]14.0 [95] 141212 121010Alloy 6061FOH all all ... [...] 22.0 [150] 16.0I [...] 16 14T1 [16.00] all [180] ... [...] [95] 16 14 all all 26.0 [180] ... [...] 16.0 [110] 16 14T42J all all 26.0 [180] ... [...] 12.0 [85] 16 14T51 [16.00] all [240] ... [...] [205] 8 7 up through 0.249 [6.30]0.250 and over [6.30] allall 38.0 [260]38.0 [260] ... [...]... [...] 35.0 [240]35.0 [240] 810 ...9Alloy 6063FOHT1K allup through 0.500 [12.50]0.501–1.000 [12.50–25.00] ... [all]allall ... [...]17.0 [115]16.0 [110] 19.0 [130]... [...]... [...] ... [...]9.0 [60]8.0 [55] 181212 [...] 161010T4, T42L up through 0.500 [12.50] all 19.0 [130] ... [...] 10.0 [70] 14 12 0.501–1.000 [12.50–25.00] all 18.0 [125] ... [...] 9.0 [60] 14 [...] 12T5 up through 0.500 [12.50] all 22.0 [150] ... [...] 16.0 [110] 8 7 0.501–1.000 [12.50–25.0] all 21.0 [145] ... [...] 15.0 [105] 8 [...] 7T52 up through 1.000 [25.00] all 22.0 [150] 30.0 [205] 16.0M [110] 8 7T6, T62L up through 0.124 [3.20] all 30.0 [205] ... [...] 25.0 [170] 8 ... 0.125–1.000 [3.20–25.00] all 30.0 [205] ... [...] 25.0 [170] 10 7Alloy 6070FT6, T62L up through 2.999 up through 32 48.0 [330] ... [...] 45.0 [310] 6 5Alloy 6351FT4T6 allup through 0.1240.125–0.749 all...... 32.0 [220]42.0 [290]42.0 [290] ... [...]... [...]... [...] 19.0 [130]37.0 [255]37.0 [255] 16810 14...9A The basis of establishment of mechanical property limits is given in Annex A1 of this specification.B To determine conformance to this specification, each value for ultimate tensile strength and for yield strength shall be rounded to the nearest 0.1 ksi [MPa] and each value for elongation to the nearest 0.5 %, both in accordance with the rounding method of Practice E29.C Elongation of full-section and sheet-type specimens is measured in 2 in.; of cut-out round specimens, in 4× specimen diameter.D For material of such dimensions that a standard test specimen cannot be taken, or for material thinner than 0.062 in., the test for elongation is not required.E Elongations in 50 mm apply for tube tested in full section and for sheet-type specimens machined from material up through 12.5 mm in thickness having parallel surfaces. Elongations in 5× diameter (5.65), where D and A are diameter and cross-sectional area of the specimen, respectively, apply to round test specimens machined from thickness over 6.30 mm. For tube of such dimensions that a standard test specimen cannot be taken, the test for elongation is not required.F These alloys are also produced in the F temper, for which no mechanical properties are specified.G Maximum tensile strength and minimum elongation apply to tubes having diameters from 1.000 in. to 4.500 in. and wall thickness from 0.050 in. to 0.169 in. only. Minimum elongation applies to tubes having diameters from 25.00 to 115.00 mm and wall thickness over 1.30 through 4.30 mm only.H Upon heat treatment, annealed (0 temper) material shall be capable of developing the mechanical properties applicable to T42 temper material, and upon solution and precipitation heat treatment shall be capable of developing the mechanical properties applicable to T62 temper material.I Yield strength is maximum [110 MPa] max.J For stress-relieved tempers (T4510, T4511, T6510 and T6511) characteristics and properties other than those specified may differ somewhat from the corresponding characteristics and properties of material in the basic temper.K Formerly designated T42 temper. Properly aged precipitation heat-treated 6063-T1 extruded products are designated T5.L While material in the T42 and T62 tempers is not available from the material producer, the properties are listed to indicate those which can usually be obtained by the user when the material is properly solution heat treated or solution and precipitation heat treated from the O (annealed) or F (as-fabricated) tempers. These properties apply when samples of material supplied in the O or F temper are heat treated by the producer to the T42 or T62 tempers to determine that the material will respond to proper thermal treatment. Properties attained by the user, however, may be lower than those listed if the material has been formed or otherwise cold or hot worked, particularly in the annealed temper, prior to solution heat treatment.M Maximum yield strength is 25.0 ksi [170 MPa].TABLE 3 Chemical CompositionA,B,CAlloy Composition, %Silicon Iron Copper Manganese Magnesium Chromium Zinc Vanadium Titanium Other ElementsD AluminumEach TotalE10603003 0.250.6 0.350.7 0.050.05–0.20 0.031.0–1.5 0.03... ...... 0.050.10 0.05... 0.03... 0.030.05 ...0.15 99.60 minFremainderAlclad 3003 3003 alloy clad inside or outside with 7072 alloy5083 0.40 0.40 0.10 0.40–1.0 4.0–4.9 0.05–0.25 0.25 ... 0.15 0.05 0.15 remainder5086 0.40 0.50 0.10 0.20–0.7 3.5–4.5 0.05–0.25 0.25 ... 0.15 0.05 0.15 remainder6061G 0.40–0.8 0.7 0.15–0.40 0.15 0.8–1.2 0.04–0.35 0.25 ... 0.15 0.05 0.15 remainder6063 0.20–0.6 0.35 0.10 0.10 0.45–0.9 0.10 0.10 ... 0.10 0.05 0.15 remainder6070 1.0–1.7 0.50 0.15–0.40 0.40–1.0 0.50–1.2 0.10 0.25 ... 0.15 0.05 0.15 remainder6351 0.7–1.3 0.50 0.10 0.40–0.8 0.40–0.8 ... 0.20 ... 0.20 0.05 0.15 remainder7072H 0.7 Si + Fe 0.10 0.10 0.10 ... 0.8–1.3 ... ... 0.05 0.15 remainderA Limits are in percent maximum unless shown as a range or stated otherwise.B Analysis shall be made for the elements for which limits are shown in this table.C For purposes of determining conformance to these limits, an observed value or a calculated value obtained from analysis shall be rounded to the nearest unit in the last right-hand place of figures used in expressing the specified limit, in accordance with the rounding method of Practice E29.D Others includes listed elements for which no specific limit is shown as well as unlisted metallic elements. The producer may analyze samples for trace elements not specified in the specification. However, such analysis is not required and may not cover all metallic Others elements. Should any analysis by the producer or the purchaser establish that an Others element exceeds the limit of Each or that the aggregate of several Others elements exceeds the limit of Total, the material shall be considered non-conforming.E Other ElementsTotal shall be the sum of unspecified metallic elements 0.010 % or more, rounded to the second decimal before determining the sum.F The aluminum content shall be calculated by subtracting from 100.00 % the sum of all metallic elements present in amounts of 0.010 % or more each, rounded to the second decimal before determining the sum.G In 1965 the requirements for Alloy 6062 were combined with those of Alloy 6061 by revision of the minimum chromium content from 0.15 to 0.04. For this reason, Alloy 6062 was cancelled.H Composition of cladding alloy as applied during the course of manufacture. The sample from finished tube shall not be required to conform to these limits.

定价: 0元 / 折扣价: 0

在线阅读 收 藏

4.1 This test method provides a standardized procedure for evaluating performance of ceramic floor tile installations under conditions similar to actual specific usages. It can be used to make comparisons between customary basic installation methods, to establish the influence of minor changes in a particular installation method, and to judge the merit of proposed novel methods.1.1 This test method covers the evaluation of ceramic floor tile installation systems, using the Robinson2-type floor tester.1.2 This test method is intended solely for evaluating complete ceramic floor tile installation systems for failure under dynamic loads and not for evaluating particular characteristics of ceramic tile, such as abrasion resistance. This test method does not claim to provide meaningful results for other than evaluating complete ceramic floor tile installation systems.1.3 The values stated in inch-pound units are to be regarded as the standard. The metric (SI) units in parentheses are for information only.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

定价: 590元 / 折扣价: 502 加购物车

在线阅读 收 藏

4.1 A petroleum products, liquid fuels, and lubricants testing laboratory plays a crucial role in product quality management and customer satisfaction. It is essential for a laboratory to provide quality data. This document provides guidance for establishing and maintaining a quality management system in a laboratory.4.1.1 The word ‘customer’ can refer to both customers internal and external to the laboratory or organization.1.1 This practice covers the establishment and maintenance of the essentials of a quality management system in laboratories engaged in the analysis of petroleum products, liquid fuels, and lubricants. It is designed to be used in conjunction with Practice D6299.NOTE 1: This practice is based on the quality management concepts and principles advocated in ANSI/ISO/ASQ Q9000 standards, ISO/IEC 17025, ASQ Manual,2 and ASTM standards such as D3244, D4182, D4621, D6299, D6300, D7372, E29, E177, E456, E548, E882, E994, E1301, E1323, STP 15D,3 and STP 1209.41.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory requirements prior to use.1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

定价: 646元 / 折扣价: 550 加购物车

在线阅读 收 藏

5.1 This practice describes procedures applicable to both shop and field conditions. More comprehensive or precise measurements of the characteristics of complete systems and their components will generally require laboratory techniques and electronic equipment such as oscilloscopes and signal generators. Substitution of these methods is not precluded where appropriate; however, their usage is not within the scope of this practice.5.2 This document does not establish system acceptance limits, nor is it intended as a comprehensive equipment specification.5.3 While several important characteristics are included, others of possible significance in some applications are not covered.5.4 Since the parameters to be evaluated and the applicable test conditions must be specified, this practice shall be prescribed only by those familiar with ultrasonic NDT technology and the required tests shall be performed either by such a qualified person or under his supervision.5.5 Implementation may require more detailed procedural instructions in the format of the using facility.5.6 In the case of evaluation of a complete system, selection of the specific tests to be made should be done cautiously; if the related parameters are not critical in the intended application, then their inclusion may be unjustified. For example, vertical linearity may be irrelevant for a go/no-go test with a flaw gate alarm, while horizontal linearity might be required only for accurate flaw-depth or thickness measurement from the display screen.5.7 No frequency of system evaluation or calibration is recommended or implied. This is the prerogative of the using parties and is dependent on application, environment, and stability of equipment.5.8 Certain sections are applicable only to instruments having receiver gain controls calibrated in decibels (dB). While these may sometimes be designated “gain,” “attenuator,” or “sensitivity” on various instruments, the term “gain controls” will be used in this practice in referring to those which specifically control instrument receiver gain but not including reject, electronic distance-amplitude compensation, or automatic gain control.5.9 These procedures can generally be applied to any combination of instrument and search unit of the commonly used types and frequencies, and to most straight-beam examination, either contact or immersed. Certain sections are also compatible with angle-beam, wheel, delay-line, and dual-search unit techniques. Their use, however, should be mutually agreed upon and so identified in the test report.5.10 The validity of the results obtained will depend on the precision of the instrument display readings. This is assumed to be ±0.04 in. (±1 mm), yielding between 1 % and 2 % of full scale (fs) readability for available instrumentation having suitable screen graticules and display sharpness.1.1 This practice describes procedures for evaluating the following performance characteristics of ultrasonic pulse-echo examination instruments and systems: Horizontal Limit and Linearity; Vertical Limit and Linearity; Resolution - Entry Surface and Far Surface; Sensitivity and Noise; Accuracy of Calibrated Gain Controls. Evaluation of these characteristics is intended to be used for comparing instruments and systems or, by periodic repetition, for detecting long-term changes in the characteristics of a given instrument or system that may be indicative of impending failure, and which, if beyond certain limits, will require corrective maintenance. Instrument characteristics measured in accordance with this practice are expressed in terms that relate to their potential usefulness for ultrasonic testing. Instrument characteristics expressed in purely electronic terms may be measured as described in Guide E1324.1.2 Ultrasonic examination systems using pulsed-wave trains and A-scan presentation (rf or video) may be evaluated.1.3 The procedures are applicable to shop or field conditions; additional electronic measurement instrumentation is not required.1.4 This practice establishes no performance limits for examination systems; if such acceptance criteria are required, these must be specified by the using parties. Where acceptance criteria are implied herein, they are for example only and are subject to more or less restrictive limits imposed by customer's and end user's controlling documents.1.5 The specific parameters to be evaluated, conditions and frequency of test, and report data required must also be determined by the user.1.6 This practice may be used for the evaluation of a complete examination system, including search unit, instrument, interconnections, fixtures and connected alarm and auxiliary devices, primarily in cases where such a system is used repetitively without change or substitution. This practice is not intended to be used as a substitute for calibration or standardization of an instrument or system to inspect any given material. There are limitations to the use of standard reference blocks for that purpose.21.7 Required test apparatus includes selected test blocks and a precision external attenuator (where specified) in addition to the instrument or system to be evaluated.1.8 Precautions relating to the applicability of the procedures and interpretation of the results are included.1.9 Alternate procedures, such as examples described in this document, or others, may only be used with customer approval.1.10 Units—The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.1.11 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.1.12 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

定价: 646元 / 折扣价: 550 加购物车

在线阅读 收 藏

5.1 Use—This practice is intended for use by parties who desire access to the national, or international, airspace as regulated by their respective CAA(s) either for a vehicle design (airworthiness) or a vehicle’s use (operational approval). In this practice, it is recognized the varying levels of complexity, need for risk assessment(s), and due diligence that should be determined in an ongoing dialogue between the CAA and the applicant. Users should consider their requirements, the purpose that the ORA is to serve, and their risk acceptance level before undertaking the ORA. Use of this practice does not preclude other initiatives or processes to identify hazardous conditions or assess and mitigate associated risks.5.2 Risk Reduced, not Eliminated—No ORA can eliminate all risk or uncertainty with regard to operations. Preparation of an ORA in accordance with this practice is intended to reduce, but may not necessarily completely eliminate, the risk of an operation in which system complexity is minimal, the operation is conducted in a lower risk environment, and the likelihood for harm to people or property, though present, is reduced to an acceptable level. As mission complexity increases, the operational environment may become less risk tolerant. For example, as the kinetic energy associated with the aircraft increases, more complex assessment/analysis tools and greater time may be required to conduct the ORA.1.1 This practice focuses on preparing operational risk assessments (ORAs) to be used for supporting small unmanned aircraft systems (sUAS) (aircraft under 55 lb (25 kg)) design, airworthiness, and subsequent operational applications to the civil aviation authority (CAA).1.2 It is expected that manufacturers and developers of larger/higher energy sUAS designs, intended to operate in controlled airspace over populated areas, will adopt many of the existing manned aircraft standards in use. These include standards such as SAE ARP4754A and ARP4761, which prescribe a “design for safety” top-down design approach to ensure the sUAS designs can reasonably meet more stringent qualitative and quantitative safety requirements. The ORA, however, remains the same for all risk profiles and will be a part of any sUAS operation.1.3 In mitigating and preventing incidents and accidents, it is understood that people generally do not seek to cause damage or injure others, and therefore, malicious acts are beyond the scope of this practice.1.4 As part of the ORA, the applicant should clearly understand and be able to articulate their intended mission for purposes of assessing safety and providing information to regulators. This documentation of a sUAS operation (mission, or set of missions) is what many refer to as a concept of operations (CONOPS).1.5 This practice is intended primarily for sUAS applicants seeking approval or certification for airworthiness or operations from their respective CAA, though sUAS manufacturers may consider this practice, along with other system safety design standards, as appropriate to identify sUAS design and operational requirements needed to mitigate hazards.1.6 Units—The values stated in inch-pound units are to be regarded as the standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

定价: 646元 / 折扣价: 550 加购物车

在线阅读 收 藏

A standard recognizes that effectiveness, safety, and durability of a RBS depends not only on the quality of the materials, but also on their proper installation.Improper installation of a RBS can reduce their thermal effectiveness, cause fire risks and other unsafe conditions, and promote deterioration of the structure in which they are installed. Specific hazards that can result from improper installation include fires caused by (1) heat buildup in recessed lighting fixtures, (2) deterioration or failure of electrical wiring components, and (3) deterioration in wood structures and paint failure due to moisture accumulation.This standard provides recommendations for the installation of radiant barrier materials in a safe and effective manner. Actual conditions in existing buildings may vary greatly and in some cases additional care should be taken to ensure safe and effective installation.This standard presents requirements that are general in nature and considered practical. They are not intended as specific recommendations. The user should consult the manufacturer for recommended application methods.1.1 This standard has been prepared for use by the designer, specifier, and installer of RBS (radiant barrier systems) for use in building construction. The scope is limited to recommendations relative to the use and installation of RBS including a surface(s) normally having a far-infrared emittance of 0.1 or less, such as metallic foil or metallic deposits unmounted or mounted on substrates. Some examples that this standard is intended to address include: (1) low emittance surfaces in vented or unvented building envelope cavities intended to retard radiant transfer across the airspace; (2) low emittance surfaces at interior building surfaces intended to retard radiant transfer to or from building inhabitants; and (3) low emittance surfaces at interior building surfaces intended to reduce radiant transfer to or from radiant heating or cooling systems. See for typical examples of use.1.2 This standard covers the installation process from pre-installation inspection through post-installation procedure. It does not cover the production of the radiant barrier materials. (See Specification C1313.)1.3 This standard is not intended to replace the manufacturer's installation instructions, but shall be used in conjunction with such instructions. This practice is not intended to supercede local, state, or federal codes.1.4 This standard assumes that the installer possesses a good working knowledge of the application codes and regulations, safety practices, tools, equipment, and methods necessary for the installation of radiant barrier materials. It also assumes that the installer understands the fundamentals of building construction that affect the installation of RBS.This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific precautionary statements, see Sections and .1.5 When the installation or use of radiant barrier materials, accessories and systems, may pose safety or health problems, the manufacturer shall provide the user appropriate current information regarding any known problems associated with the recommended use of the company's products and shall also recommend protective measures to be employed in their safe utilization. The user shall establish appropriate safety and health practices and determine the applicability of regulatory requirements prior to use.

定价: 0元 / 折扣价: 0

在线阅读 收 藏

4.1 The loss of sterile barrier system integrity may occur as a result of physical properties of the materials and adhesive or cohesive bonds degrading over time or by subsequent dynamic events during shipping and handling, or both. Accelerated and real time aging verifies the time-related aspects of potential integrity loss only.4.2 ANSI/AAMI/ISO 11607–1: 2019, sub-clause 6.1.3, states that “the packaging system shall provide physical protection in order to maintain integrity of the sterile barrier system.” Sub-clause 6.1.6 states that, “A terminally sterilized sterile barrier system with its protective packaging, if included, shall be designed to, maintain sterility through exposure to expected conditions and hazards during the specified processing, storage, handling, and distribution until that SBS is opened at the point of use or until the expiry date.” Sub-clause 8.3.1 states, “Stability testing shall demonstrate that the sterile barrier system maintains integrity over time.” Sub-clause 8.3.3 states, “Stability testing, using accelerated aging protocols, shall be regarded as sufficient evidence for claimed expiry dates until data from real-time aging studies are available.”4.3 Real time aging programs provide the best data to ensure that sterile barrier system/medical device materials and sterile barrier system/medical device integrity do not degrade over time. However, due to market conditions in which products may become obsolete in a short time, and the desire to get new products to market in the shortest possible time, real time aging studies do not meet this objective. Accelerated aging studies can provide an alternative means of screening for possible aging-related failure mechanisms in the SBS or medical device. To ensure that accelerated aging studies represent real time effects, real time aging studies must be conducted in parallel to accelerated studies. Real time studies must be carried out to the claimed shelf life of the product and be performed to their completion.4.4 Conservative accelerated aging factors (AAFs) must be used if little is known about the sterile barrier system material being evaluated. More aggressive AAFs may be used with documented evidence to show a correlation between real time and accelerated aging.4.5 When conducting accelerated aging programs for establishing expiry dating claims, it must be recognized that the data obtained from the study is based on conditions that simulate the effects of aging on the materials. The resulting creation of an expiration date or shelf life is based on the use of a conservative estimate of the aging factor (that is, Q10) and is tentative until the results of real time aging studies are completed on the sterile barrier system.NOTE 1: Determining AAFs are beyond the scope of this guide.61.1 This guide provides information for developing accelerated aging protocols to model the possible effects of the passage of time on the sterile integrity of the sterile barrier system (SBS), as defined in ANSI/AAMI/ISO 11607–1: 2019 and the physical properties of their component packaging materials. Guidance for developing accelerated aging protocols may also be used for medical devices and medical device materials.1.2 Information obtained using this guide may be regarded as sufficient evidence for expiration date claims for medical devices and sterile barrier systems until data from real-time aging studies are available.1.3 The accelerated aging guideline addresses sterile barrier systems as a whole with or without devices. The sterile barrier system material and device interaction compatibility that may be required for new product development or the resulting evaluation is not addressed in this guide.1.4 Real-time aging protocols are not addressed in this guide; however, it is essential that real-time aging studies be performed to confirm the accelerated aging test results using the same methods of evaluation. Real-time aging (stability) is the requirement of ANSI/AAMI/ISO 11607–1: 2019.1.5 Methods used for sterile barrier system performance validation, which include, environmental challenge, distribution, handling, and shipping events, are used for package performance (event-related loss of integrity) testing and are beyond the scope of this guide.1.6 This guide does not address environmental challenging that simulates extreme climactic conditions that may exist in the shipping and handling environment. Refer to Practice D4332 for standard conditions that may be used to challenge the sterile barrier system to realistic extremes in temperature and humidity conditions. See Terminology F17 for a definition of “environmental challenging.”1.7 The data obtained from accelerated aging studies is not to be used as a manner of establishing label storage conditions for sterile barrier systems.1.8 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.1.10 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

定价: 590元 / 折扣价: 502 加购物车

在线阅读 收 藏

5.1 A major factor affecting the life of insulating materials is thermal degradation. It is possible that other factors, such as moisture and vibration, will cause failures after the material has been weakened by thermal degradation.5.2 Electrical insulation is effective in electrical equipment only as long as it retains its physical and electrical integrity. The following are potential indicators of thermal degradation: weight change, porosity, crazing, and generally a reduction in flexibility. Thermal degradation is usually accompanied by an ultimate reduction in dielectric breakdown.5.3 This test method is useful in determining the thermal endurance of coating powders applied over a copper or aluminum substrate material.1.1 This test method provides a procedure for evaluating thermal endurance of coating powders by determining the length of aging time at selected elevated temperatures required to achieve dielectric breakdown at room temperature at a pre-determined proof voltage. Thermal endurance is expressed in terms of a temperature index.1.2 This test method is applicable to insulating powders used over a substrate material of copper or aluminum.1.3 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system are not necessarily exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems is likely to result in non-conformance with the standard.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 7.1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

定价: 590元 / 折扣价: 502 加购物车

在线阅读 收 藏

5.1 A major factor affecting the life of insulating materials is thermal degradation. It is possible that other factors, such as moisture and vibration, will cause failures after the material has been weakened by thermal degradation.5.2 Electrical insulation is effective in electrical equipment only as long as it retains its physical and electrical integrity. The following are potential indicators of thermal degradation: weight change, porosity, crazing, and generally a reduction in flexibility. Thermal degradation is usually accompanied by an ultimate reduction in dielectric breakdown.5.3 This test method is useful in determining the thermal endurance of coating powders applied over a steel substrate material.1.1 This test method provides a procedure for evaluating thermal endurance of coating powders by determining the length of aging time at selected elevated temperatures required to achieve dielectric breakdown at room temperature at a pre-determined proof voltage. Thermal endurance is expressed in terms of a temperature index.1.2 This test method is applicable to insulating powders used over a substrate material of steel.1.3 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system are not necessarily exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems is likely to result in non-conformance with the standard.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 7.1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

定价: 590元 / 折扣价: 502 加购物车

在线阅读 收 藏

This specification covers requirements for multiplayer pipe type 2 and compression fittings for hot and cold drinking water systems. Multilayer pipe type 2 is produced using a butt-welded aluminum pipe as a core, with an extruded inside layer of crosslinked polyethylene (PEX). An adhesive layer is used to bond the inside layer to the wall of the aluminum pipe. An outer layer of polyethylene (PE) and an adhesive layer are extruded to the outer wall of the aluminum pipe. This specification includes compression fittings and thread or solder adapters for use with pipe and fittings. The pipe dimensions, compression-fitting dimensions, burst pressure, thermal cycling, and excessive temperature-pressure capability shall be in conformance to the specification.1.1 This specification covers requirements for multilayer pipe type 2 and compression fittings for hot and cold drinking-water systems, with a maximum pressure rating of 1000 kPa (145 psi) at 82°C (180°F).Note 1—Multilayer Pipe Type 2Construction-based pressure rated pipe comprising more than one layer in which at least 60 % of the wall thickness is polymeric material.1.2 Multilayer pipe type 2 is produced using a butt-welded aluminum pipe as a core, with an extruded inside layer of crosslinked polyethylene (PEX). An adhesive layer is used to bond the inside layer to the wall of the aluminum pipe. An outer layer of polyethylene (PE) and an adhesive layer are extruded to the outer wall of the aluminum pipe.1.3 Multilayer pipe type 2 is produced in configurations 1 and 2, as referenced in Fig. 1.1.4 This specification includes compression fittings, which are referenced in Fig. 2 .1.5 Specifications for thread or solder adapters for use with pipe and fittings meeting the requirements of this specification are given in Annex A1 and Annex A2.1.6 The following precautionary caveat pertains only to the test method portion of this specification: This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.1.7 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.

定价: 0元 / 折扣价: 0

在线阅读 收 藏

This specification defines the masses to be used when testing rescue systems and components. The masses represent personnel and equipment that may be attached to a rescue system or components. However, the masses do not represent any particular type or kind of rescuer or equipment. The masses shall be classified as follows: Type I; Type II; Type III; Type IV; and Type V.1.1 This specification defines the masses to be used when testing rescue systems and components.1.2 The masses represent personnel and equipment that may be attached to a rescue system or components. However, the masses do not represent any particular type or kind of rescuer or equipment.1.2.1 The masses chosen have been used in the past or are in current use in testing of rescue systems and components. Limiting testing to the masses listed in this specification allows meaningful comparisons between past, current, and future test results.1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.4 The user of this specification shall determine which mass(es) represent(s) the personnel and equipment attached to the system or component under test.1.5 For the purposes of this specification, mass and weight are synonymous when the object(s) representing the mass(es) are weighed in air anywhere on Earth.1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory requirements prior to use.

定价: 0元 / 折扣价: 0

在线阅读 收 藏
535 条记录,每页 15 条,当前第 1 / 36 页 第一页 | 上一页 | 下一页 | 最末页  |     转到第   页